Feasibility of surface guided radiotherapy for patient positioning in breast radiotherapy versus conventional tattoo-based setups- a systematic review

Open AccessPublished:April 16, 2022DOI:https://doi.org/10.1016/j.tipsro.2022.03.001

      Abstract

      Background

      Traditionally tattoos are used for patient setup in radiotherapy. However they may pose challenges for the radiotherapists to achieve precise patient alignment, and serve as a permanent visual reminder of the patient’s diagnosis and often challenging cancer journey. The psychological impact of tattoos has been recognized in recent years. The increasing complexity of treatment techniques and the utilization of hypofractionated regimes, requires an enhanced level of accuracy and safety. Surface guided radiotherapy (SGRT) enables improvements in the accuracy and reproducibility of patient isocentric and postural alignment, enhanced efficiency, and safety in breast radiotherapy.

      Purpose

      The aim of this review was to compare the accuracy and reproducibility of SGRT to conventional tattoo-based setups in free-breathing breast radiotherapy and to determine if SGRT can reduce the frequency of routine image guided radiotherapy (IGRT).

      Materials and Methods

      A systematic literature review was performed as per PRISMA guidelines. Papers identified through PubMed, Embase, Web of Science and Google Scholar database searches between 2010 and 2021, were critically appraised. Systematic, random, mean residual errors and 3D vector shifts as determined by IGRT verification were analysed.

      Results

      A review of 13 full papers suggests SGRT improves the accuracy and reproducibility of patient setup in breast radiotherapy with consistent reductions in the residual errors. There appears to be a good correlation between SGRT setups and radiographic imaging. The frequency of IGRT and the corresponding dose could potentially be reduced. Additionally, SGRT improves treatment efficiency.

      Conclusion

      SGRT appears to have improved the accuracy and reproducibility of patient setup and treatment efficiency of breast radiotherapy compared to conventional tattoo/laser-based method, with the potential to reduce the frequency of routine IGRT. The reliance on tattoos in breast radiotherapy are likely to become obsolete with positive implications for both patients and clinical practice.

      Keywords

      Introduction

      Breast cancer and radiotherapy

      Breast cancer is the most common form of cancer in the UK with approximately 55,000 women diagnosed each year. The 5-year and 10-year survival rates are 86% and 78% respectively, which can be attributed to improvements in screening, diagnosis, and treatment [

      Breast Cancer Now. Breast Cancer- The numbers. 2015. Available from: https://breastcancernow.org/sites/default/files/files/breast-cancer-stats-sheet-february-2015.pdf.

      ]. Adjuvant radiotherapy is commonly administered to the chestwall, whole breast or partial breast to reduce the risk of local recurrence [

      National Institute for Health and Care Excellence (NICE). Early and locally advanced breast cancer: diagnosis and management- NICE guideline [NG101]. Available from: https://www.nice.org.uk/guidance/ng101/chapter/Recommendations#radiotherapy.

      ].
      In the UK, 40 Gy in 15 fractions is the standard of care for breast radiotherapy [
      • Koulis T.A.
      • Phan T.
      • Olivotto I.A.
      Hypofractionated whole breast radiotherapy: current perspectives.
      ]. Recently the results of the FAST_Forward trial validated the use of 26 Gy in 5 consecutive fractions for early-stage breast cancer, post breast conservation surgery [
      • Murray Brunt A.
      • et al.
      Hypofractionated breast radiotherapy for 1 week versus 3 weeks (FAST-Forward): 5-year efficacy and late normal tissue effects results from a multicentre, non-inferiority, randomised, phase 3 trial.
      ]. These hypofractionated schedules have demonstrated comparable 5-year local tumour control rates, and acute and late normal tissue effects, with the advantage of reduced overall treatment time and patient convenience [
      • Murray Brunt A.
      • et al.
      Hypofractionated breast radiotherapy for 1 week versus 3 weeks (FAST-Forward): 5-year efficacy and late normal tissue effects results from a multicentre, non-inferiority, randomised, phase 3 trial.
      ,
      • De Felice F.
      • et al.
      Relation between Hypofractionated Radiotherapy, Toxicity and Outcome in Early Breast Cancer.
      ]. The impact of the COVID19 pandemic on radiotherapy services reaffirms the benefits of these fractionation regimes [
      • Coles C.E.
      • et al.
      International Guidelines on Radiation Therapy for Breast Cancer During the COVID-19 Pandemic.
      ]. Hypofractionation requires an additional level of safety due to the higher fractional dose, tighter margins, and steeper dose gradients than conventional fractionation. To ensure precise dose delivery to the target and sparing of the surrounding heart and lung, interfractional and intrafractional uncertainties must be minimised [
      • Laine A.M.
      • et al.
      The Role of Hypofractionated Radiation Therapy with Photons, Protons, and Heavy Ions for Treating Extracranial Lesions.
      ].
      Tangential whole breast radiotherapy is generally robust to systematic and random setup errors owing to larger margins; however, setup accuracy and target positioning are crucial in volumetric arc therapy (VMAT) and accelerated partial breast irradiation (APBI), due to the greater impact of interfraction and intrafraction discrepancies on the delivered dose [
      • Michalski A.
      • et al.
      Inter- and intra-fraction motion during radiation therapy to the whole breast in the supine position: a systematic review.
      ].
      Typically, an anterior and two lateral tattoos are used as the primary reference for initial patient alignment to ensure reproducibility to the scanned position. In some clinics, additional markings are used to indicate the isocentre or field borders [
      • Townend C.
      • et al.
      A review of permanent marking for radiotherapy in the UK.
      ]. The quality of the delivered treatments is greatly dependent on the ability to position the patient accurately and reproducibly throughout the course of radiotherapy [
      • Pham K.
      • et al.
      Comparison of surface guided radiation therapy and tattooing techniques for breast radiotherapy.
      ].
      Image guided radiotherapy (IGRT) is performed prior to treatment delivery, typically using a bone and soft tissue registration to localise the target and correct for interfraction setup deviations [
      • Hoisak J.D.P.
      • Pawlicki T.
      The Role of Optical Surface Imaging Systems in Radiation Therapy.
      ,
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ]. The tumour bed may be demarcated with surgical clips which aid in the delineation of the target volume and may be used for image registration during IGRT. The frequency of IGRT varies per department, ranging from daily, or the first three fractions and thereafter once weekly [
      • Nabavizadeh N.
      • et al.
      Image Guided Radiation Therapy (IGRT) Practice Patterns and IGRT's Impact on Workflow and Treatment Planning: Results From a National Survey of American Society for Radiation Oncology Members.
      ].

      Challenges associated with breast radiotherapy

      Tattoo-based setups pose some challenges, namely, mobility of the skin and corresponding variability of the tattoo positions. Radiation therapists often need to manipulate the skin to ensure correct alignment of the lasers to the tattoo markings [
      • Leong B.
      • Padilla L.
      Impact of use of optical surface imaging on initial patient setup for stereotactic body radiotherapy treatments.
      ]. Identification of tattoos against other skin markings, misidentification of current tattoos and those from previous irradiation, or ease of visualisation on darker skin tone patients, proves challenging for therapists and increases the probability of setup errors [
      • Leong B.
      • Padilla L.
      Impact of use of optical surface imaging on initial patient setup for stereotactic body radiotherapy treatments.
      ,
      • Landeg S.J.
      • et al.
      A randomized control trial evaluating fluorescent ink versus dark ink tattoos for breast radiotherapy.
      ,
      • Padilla L.
      • et al.
      Assessment of interfractional variation of the breast surface following conventional patient positioning for whole-breast radiotherapy.
      ].
      Today, there is greater awareness around the cosmetic and psychological impact of tattoos. Some patients may be impartial to these however for others, tattoos may serve as a visible permanent reminder of their challenging and sometimes distressing cancer journey [
      • Rigley J.
      • Robertson P.
      • Scattergood L.
      Radiotherapy without tattoos: Could this work?.
      ]. A group of patients who underwent breast radiotherapy with the use of tattoos or skin marks were surveyed in the USA, indicated they were willing to invest additional resources (travel, time, and money) to avoid tattoos/skin marks [
      • Moser T.
      • et al.
      Radiotherapy tattoos: Women's skin as a carrier of personal memory—What do we cause by tattooing our patients?.
      ]. These findings are increasingly more pertinent and should be explored deeper, especially in an era where technological advances have enabled suitable alternatives.
      More importantly tattoos do not provide postural information of the arm and chin, which can impact reproducibility and ultimately the target positioning [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ,
      • Batista V.
      • et al.
      Clinical paradigms and challenges in surface guided radiation therapy: Where do we go from here?.
      ]. Ultraviolet (UV)-visible tattoos as a replacement for dark ink tattoos has been explored however it requires additional time locate these during patient setup [
      • Townend C.
      • et al.
      A review of permanent marking for radiotherapy in the UK.
      ,
      • Landeg S.J.
      • et al.
      A randomized control trial evaluating fluorescent ink versus dark ink tattoos for breast radiotherapy.
      ,
      • Moser T.
      • et al.
      Radiotherapy tattoos: Women's skin as a carrier of personal memory—What do we cause by tattooing our patients?.
      ].
      Breast tissue is highly deformable and image registration (MV or KV) on bony anatomy alone may result in larger interfraction variation and compromised target coverage [
      • Padilla L.
      • et al.
      Assessment of interfractional variation of the breast surface following conventional patient positioning for whole-breast radiotherapy.
      ]. Setup reproducibility is often challenging in patients with higher body mass index or larger pendulous breast [
      • Probst H.
      • et al.
      A systematic review of methods to immobilise breast tissue during adjuvant breast irradiation.
      ]. Breast tissue is susceptible to contour changes, which could negatively impact the planned dose distribution. SGRT may address these challenges and provide suitable solution for reproducible breast alignment and great confidence with partial breast irradiation [
      • Batista V.
      • et al.
      Clinical paradigms and challenges in surface guided radiation therapy: Where do we go from here?.
      ,
      • Hattel S.H.
      • et al.
      Evaluation of setup and intrafraction motion for surface guided whole-breast cancer radiotherapy.
      ,
      • Chang A.J.
      • et al.
      Video surface image guidance for external beam partial breast irradiation.
      ].

      Surface guided radiotherapy

      SGRT is an optical surface tracking system that analyses the patient surface topography and is sensitive to contour and anatomical variations. These discrepancies are more apparent to the radiation therapist than conventional setup methods and provide quantitative information to aid in the clinical decision-making process [
      • Padilla L.
      • et al.
      Assessment of interfractional variation of the breast surface following conventional patient positioning for whole-breast radiotherapy.
      ,
      • Alderliesten T.
      • et al.
      Breast-shape changes during radiation therapy after breast-conserving surgery.
      ]. In an era of SGRT, the relevance and reliance on tattoos for accurate patient setup may be diminished [
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ,
      • Stanley D.N.
      • et al.
      Comparison of initial patient setup accuracy between surface imaging and three point localization: A retrospective analysis.
      ]. Although omitting tattoos seems a minute change, it is a paradigm shift as radiation therapists have relied on these markings for decades. Advances in technology, can facilitate improvements in the patient experience and satisfaction whilst maintaining a high degree of accuracy [
      • Batista V.
      • et al.
      Clinical paradigms and challenges in surface guided radiation therapy: Where do we go from here?.
      ,
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ].
      SGRT uses non-invasive, non-ionising optical technology to generate a 3D surface of the patient which is matched to a reference surface either from CT or acquired with the SGRT cameras itself. A rigid or deformable surface registration algorithm computes the differences between the live and reference position, providing real-time feedback/displacements in 6 degrees of freedom (6DOF). Surface registration is based on a region of interest (ROI) (potentially user-defined) as opposed to 3 tattoos or skin marks alone. [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ,
      • Stanley D.N.
      • et al.
      Comparison of initial patient setup accuracy between surface imaging and three point localization: A retrospective analysis.
      ,
      • Freislederer P.
      • et al.
      Recent advanced in Surface Guided Radiation Therapy.
      ].
      During traditional tattoo-based setups, the patient is aligned such that the lasers are coincident with the reference marks. The patient is manually or automatically shifted to the planned isocentre position. Alternatively, isocentre marks enables direct alignment to the isocentre (Fig. 1-Process A). The addition of SGRT, allows for fine-tuning of the isocentric alignment in 6DOF including enhanced postural correction. SGRT does not require any visible skin markings. The patient is clinically straightened and shifted directly to the approximated anatomical location. Patient adjustments are based on the real-time displacement information provided by the SGRT system (Fig. 1-Process B). Redundancies in the tattoo/laser-based setup process may improve efficiency [
      • Hoisak J.D.P.
      • Pawlicki T.
      The Role of Optical Surface Imaging Systems in Radiation Therapy.
      ].
      Figure thumbnail gr1
      Fig. 1Flowcharts depicting the steps in a typical breast radiotherapy patient setup using conventional tattoos and lasers (process A) vs SGRT (process B).
      Restricted shoulder mobility and associated discomfort post-surgery may hinder setup reproducibility or result in interfraction or intrafractional discrepancies. The larger field of view (FOV) with SGRT allows for accurate alignment of the patient arm, chin, breast tissue, and surrounding chestwall [
      • Hoisak J.D.P.
      • Pawlicki T.
      The Role of Optical Surface Imaging Systems in Radiation Therapy.
      ,
      • Padilla L.
      • et al.
      Assessment of interfractional variation of the breast surface following conventional patient positioning for whole-breast radiotherapy.
      ,

      Paxton JDPHAB, Waghorn B, Pawlicki T. Surface Guided Radiotherapy. 2020, Florida: CRC Press.

      ].
      Many studies have highlighted SGRT as a convenient and effective tool in improving the accuracy and efficiency of initial patient setup, patient comfort, safety, and clinical outcomes in breast cancer treatments [
      • Pham K.
      • et al.
      Comparison of surface guided radiation therapy and tattooing techniques for breast radiotherapy.
      ,
      • Batista V.
      • et al.
      Clinical paradigms and challenges in surface guided radiation therapy: Where do we go from here?.
      ,
      • Stanley D.N.
      • et al.
      Comparison of initial patient setup accuracy between surface imaging and three point localization: A retrospective analysis.
      ]. SGRT is gaining acceptance and adoption and has provided a viable option for tattoo-free patient setup [
      • Rigley J.
      • Robertson P.
      • Scattergood L.
      Radiotherapy without tattoos: Could this work?.
      ,
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ,
      • Stanley D.N.
      • et al.
      Comparison of initial patient setup accuracy between surface imaging and three point localization: A retrospective analysis.
      ]. Where a good correlation exists between the surface and internal anatomy, SGRT offers the potential to reduce the frequency of radiographic imaging and the associated patient imaging dose and may be a suitable alternative on non-imaging days [

      Bekke SN. Evaluation of Optical Surface Scanning of Breast Cancer Patients for Improved Radiotherapy, in DTU Nutech. 2018. p. 175.

      ,
      • Liu M.
      • et al.
      Application of Optical Laser 3D Surface imaging system (Sentinel) in breast cancer radiotherapy.
      ,
      • Wei X.
      • et al.
      Setup errors and effectiveness of Optical Laser 3D Surface imaging system (Sentinel) in postoperative radiotherapy of breast cancer.
      ]. Real-time non-ionising feedback ensures intrafraction motion can be identified, quantified and promptly controlled. Additionally, radiation delivery can be automatically interrupted if a predefined threshold is exceeded [
      • Freislederer P.
      • et al.
      Recent advanced in Surface Guided Radiation Therapy.
      ].
      The aim of this research was to compare the accuracy and reproducibility of SGRT to conventional tattoo-based setups in free-breathing breast radiotherapy and to determine if SGRT can reduce the frequency of routine IGRT. This review focused on SGRT for initial patient positioning in breast radiotherapy compared to conventional tattoo/skin marks and laser-based setups to assess the impact on accuracy, reproducibility and correlation to IGRT.

      Materials and methods

      A systematic review of the literature was conducted using the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) [

      PRISMA Transparent reporting of systematic reviews and meta-analyses. Welcome to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) website. 2021. Available from: http://www.prisma-statement.org/.

      ] (Fig. 2).
      Figure thumbnail gr2
      Fig. 2PRISMA flow diagram of information through the phases of the systematic review and corresponding results of the database search, screening and exclusion.
      Supplementary Table 1 provides the search strategies for all databases. A total of 20 papers were identified in PubMed, 69 in Embase and 42 in Web of Science. A Google scholar search on SGRT and breast radiotherapy limited to 2010–2020 resulted in a further 132 articles.
      Breast cancer is most prevalent in females hence this review focused on clinical studies of women receiving breast radiotherapy in free breathing for any disease stage. Left and right-sided whole breast or chestwall were included. All external beam radiotherapy delivery techniques were considered including proton beam irradiation.
      Papers needed to report interfractional setup errors based on IGRT analysis and the initial patient setup performed with SGRT and their standard setup procedure (tattoo/skin marks). In studies exploring tattoo-less radiotherapy, SGRT for initial patient alignment was mandatory. Studies analysing intrafraction motion management with SGRT were included if the authors reported setup error data for both setup methods. Phantom studies or those conducted using inhouse developed optical surface tracking systems were excluded. SGRT systems needed to be commercially available with the relevant FDA clearance or CE marking. The full inclusion and exclusion criteria are provided in Supplementary Table 2.
      Papers that that met the inclusion criteria, were selected for further screening. These papers were appraised in detail, including their reference list for other relevant articles. The effective public health practice project (EPHPP) quality assessment tool for quantitative studies was used to assess the quality of the literature included in the review and provided an overall rating. Various components were assessed and scored which included study design, data collection and analyses to name a few. A global rating was there decided per paper with a score of 1–3 indicating strong, moderate and weak quality [

      McMaster University School of Nursing. EPHPP- Effective Public Health Practice Project. 2018. Available from: https://merst.ca/wp-content/uploads/2018/02/quality-assessment-tool_2010.pdf.

      ]. A summary of the reviewed articles is given in Supplementary Table 3.
      Systematic and random translational and rotational setup errors for SGRT and conventional tattoo setups were extracted. Where reported, the mean residual errors and 3D vector displacements were obtained from the included papers. IGRT (MV/KV planar imaging or CBCT) is the current clinical gold standard for patient setup verification. Following initial patient alignment, the acquired radiographic images are registered to the planning reference image and these displacements constitute the residual setup errors. Image registration criteria varies per department however generally based on bone or bone and soft tissue. The group mean systematic and random setup errors are defined by Batin et al [
      • Batin E.
      • et al.
      Can surface imaging improve the patient setup for proton postmastectomy chest wall irradiation?.
      ] as the mean of all the individual mean patient errors and average of standard deviation (SD) of all the individual patients means, respectively.

      Results

      This review analysed tattoo and SGRT-based patient positioning in terms of residual setup errors and the correlation of SGRT to IGRT. 16 full papers were assessed and 3 were excluded.
      The papers reviewed encompassed breast or chestwall treated with tangential technique, locoregionally, or partial breast irradiation with photon or proton therapy. The study population in 10/13 papers was in excess of 20 patients (range: 10–76). However, Stanley et al [
      • Stanley D.N.
      • et al.
      Comparison of initial patient setup accuracy between surface imaging and three point localization: A retrospective analysis.
      ] did not report the number of patients but rather a total of 6000 fractions with 600–900 per tumour site for each setup method.
      A summary of the included studies is given in Supplementary Table 3. The imaging modality and registration method used varied between studies as per departmental protocol; however, in nearly 50% of all studies, orthogonal KV imaging was used. Rotational values were not provided in all papers.
      Generally, 3 skin markings (range: 2–6) were used as a reference for initial patient alignment. The radiation therapists visually verified that the laser coincided with the anterior and lateral tattoos. For SGRT setups, the displacements based on the surface registration of the current patient position against the DICOM or SGRT reference surface, were reduced as close as possible to zero. The general observed SGRT tolerances were 2–3 mm and 2-3degrees for translations and rotations, respectively.
      The overall observations of this review indicate SGRT setups were more accurate compared to the traditional tattoo/laser-based methods with respect to the systematic, random errors and the 3D vector shifts. Mean systematic and random errors in the vertical, longitudinal, and lateral directions were larger for tattoo setups. SGRT patient alignment resulted in a consistent reduction in the overall 3D vector shifts.

      Systematic errors

      The SGRT mean systematic errors in the 3 translational directions were reduced when compared to the tattoo setups alone (Fig. 3). The mean systematic translational errors ranged between 2.2 and 4.4 mm for tattoo and laser-based setups and were constrained to within 0.8–2.9 mm with SGRT. In the vertical and lateral directions, average reductions in the residual errors were in the region of 50% (1–2.5 mm) and 40% (0.9–1.5 mm) respectively with SGRT. Greater variability in the percentage difference between both modalities was seen in the longitudinal direction. The systematic rotational errors were comparable for both tattoos and SGRT setups and typically below 1 degree (range: 0.4–1.2°).
      Figure thumbnail gr3
      Fig. 3Mean systematic errors are displayed for both setup modalities (tattoo/laser and SGRT). Mean is representative of the mean of all the individual patient mean errors in the 3 translational directions (VERT- vertical, LNG- longitudinal, LAT-lateral). A reduction in the mean systematic errors in the vertical, longitudinal, and lateral directions are seen when SGRT is utilised (indicated by the striped bars). The treatment sites include whole breast, partial breast irradiation, and proton beam chest wall irradiation. Significant improvements are noted in the lateral direction for Cravo et al and all 3 directions for Batin et al.

      Random errors

      Similarly, reductions in the magnitude of the random translational errors were seen in the vertical, longitudinal, and lateral directions with the utilisation of SGRT. The percentage change between both methods ranged between 13 - 63% (0.4–1.8 mm) (Fig. 4). Jimenez et al [
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ] reported only translational errors. The random rotational errors were comparable for both setup methods (range: 0.6–1.5°).
      Figure thumbnail gr4
      Fig. 4The histogram indicates the mean translational random errors for tattoo (solid) and SGRT setups (striped bars). Random error is defined as the mean or root mean square (RMS) of the standard deviation (SD) of all the individual patients' means. Similar to the mean systematic errors, there is a corresponding reduction in the magnitude of the random errors in all 3 directions when SGRT is used to correct the initial patient position.
      Substantial improvements were seen by Batin et al [
      • Batin E.
      • et al.
      Can surface imaging improve the patient setup for proton postmastectomy chest wall irradiation?.
      ] for SGRT systematic and random translational and rotational errors. Residual errors with tattoo alignment were reduced from 3.2 ± 2.0 mm, 3.1 ± 3.0 mm, 2.6 ± 2.5 mm in the vertical, longitudinal and lateral directions to 1.3 ± 1.3 mm, 0.8 ± 1.2 mm and 1.5 ± 1.4 mm with SGRT.

      Mean residual setup errors

      The mean positional discrepancies of tattoo-only alignment were larger than with SGRT (Fig. 5 and Supplementary Table 4). However, 2 studies demonstrated a larger vertical error for SGRT setups. Jimenez et al [
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ] observed a non-statistically significant reduction in systematic and random errors with SGRT however the mean vertical error was larger with SGRT (2.3 vs −1.4 mm). Kügele et al [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ] also reported a larger mean vertical error in the SGRT breast tangent field group (1.5 ± 1.7 mm vs 0.6 ± 3.7 mm, p < 0.01). Tattoo setups demonstrated a larger variability in setup errors in all 3 translational directions. The SD with SGRT was consistently smaller than that of the tattoo-only group demonstrating a higher degree of consistency. In some cases, the magnitude of the errors was reduced however the shifts were in opposite directions.
      Figure thumbnail gr5
      Fig. 5Group mean residual setup errors and its SD for the studies reviewed. The histogram represents the mean of all the mean individual patients’ errors. The SD deviation is indicated by the error bars. SD is not reported by Liu et al. The mean patient positional discrepancies are reduced for SGRT setups however in Jimenez et al and Kügele et al (tangent cohort) an improvement in the vertical direction is seen with tattoo/laser-based setups. The distribution and variability of the shifts are consistently lower for SGRT assisted setup. The methodology and analysis used by Liu et al varies from other studies which explains the directional variations despite a reduction in the mean residual setup errors with SGRT.

      3D vector displacements

      The 3D vector shifts of the SGRT translational setup errors were consistently smaller than that of the tattoo setups (Fig. 6). Statistically significant differences were noted for Kügele et al [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ] (Tangents and locoregional groups respectively) (4.2 vs 2.4 mm and 4.7 vs 4 mm, p < 0.01), Stanley et al [
      • Stanley D.N.
      • et al.
      Comparison of initial patient setup accuracy between surface imaging and three point localization: A retrospective analysis.
      ] (14 ± 7 mm vs 6 ± 2 mm, p < 0.01), Chang et al [
      • Chang A.J.
      • et al.
      Video surface image guidance for external beam partial breast irradiation.
      ] (8.8 ± 4.2 mm vs 4 ± 2.3 mm, p = 0.02), Kost et al [
      • Kost S.
      • et al.
      Assessment of Setup Accuracy Using Anatomical Landmarks for Breast and Chest Wall Irradiation With Surface Guided Radiation Therapy.
      ] (Tangents and locoregional- skin mean groups respectively) (2.9 ± 1.3 mm vs 1.8 ± 1.0 mm and 3.5 ± 1.9 mm vs 2.5 ± 1.4 mm, p < 0.001). Lateral setup errors reported by Hattel et al [
      • Hattel S.H.
      • et al.
      Evaluation of setup and intrafraction motion for surface guided whole-breast cancer radiotherapy.
      ] were significantly improved (p = 0.0009) with SGRT however the contrary was noted in the vertical direction (p = 0.00004).
      Figure thumbnail gr6
      Fig. 63D vector shifts for free-breathing breast radiotherapy patients. A reduction in the average magnitude of the 3D shift corrections were seen with SGRT setups, suggesting improved setup accuracy. Residual setup errors for tattoos were in the range of 2.4–14 mm and 1.8–6 mm with SGRT. Statistically significant differences were noted for Kügele et al (Tangents and locoregional groups) p < 0.01, Stanley et al (p < 0.01), Chang et al (p = 0.02, <0.05)), Kost et al (Tangents- skin mean) p < 0.001) and in lateral direction for Hattel et al (P = 0.0009).
      Reductions in the magnitude of the SGRT 3D vector shifts were observed by Jimenez et al [
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ], Rigley et al [
      • Rigley J.
      • Robertson P.
      • Scattergood L.
      Radiotherapy without tattoos: Could this work?.
      ], and Kost et al [
      • Kost S.
      • et al.
      Assessment of Setup Accuracy Using Anatomical Landmarks for Breast and Chest Wall Irradiation With Surface Guided Radiation Therapy.
      ] (tangent and locoregional chestwall mean errors) however these were not statistically significant.
      Overall magnitude of the residual setup uncertainties for SGRT and tattoos were in the range of 1.8–6 mm and 2.4–14 mm respectively, with 91% of reported SGRT errors below 5 mm (Fig. 6 and Supplementary Table 5).

      SGRT and IGRT correlation

      The inclusion of SGRT, improved the setup accuracy compared to tattoo/laser-based setups or was at least of similar magnitude, as verified by IGRT. Chang et al [
      • Chang A.J.
      • et al.
      Video surface image guidance for external beam partial breast irradiation.
      ] demonstrated, a substantial reduction of the mean residual setup errors in the vertical, longitudinal and lateral directions with SGRT alignment (1.9 ± 2.2, 1.8 ± 1.9, 1.8 ± 2.1 mm) compared to tattoo setups (3.2 ± 2.9, 4.2 ± 3.5, 4.7 ± 5.3 mm) analysed on orthogonal KV imaging (Fig. 7). A corresponding statistically significant reduction in the SGRT 3D vector was seen (4.0 ± 2.3 mm vs 8.3 ± 3.8 mm, p = 0.02). The distribution of the errors with tattoo setups indicates a higher degree of variability. Ma et al [
      • Ma Z.
      • et al.
      Optical Surface Management System for Patient Positioning in Interfractional Breast Cancer Radiotherapy.
      ] suggested a good correlation between SGRT and IGRT with comparable mean setup errors and standard deviations. Similarly, Deantonio et al [
      • Deantonio L.
      • et al.
      Detection of setup uncertainties with 3D surface registration system for conformal radiotherapy of breast cancer.
      ], showed non-significant systematic error in the longitudinal (p = 0.69) and lateral (p = 0.67) directions for both setup methods. An improvement in the vertical random error was noted in favour of IGRT (1.2 ± 0.4 mm vs 1.6 ± 0.6 mm).
      Figure thumbnail gr7
      Fig. 7Group mean residual translational errors + SD for SGRT and IGRT. A good correlation of SGRT to IGRT is seen in Ma et al with comparable residual setup errors and SD for both methods. Similarly, in Deantonio et al, a paired t-test showed non-significant systematic error in the longitudinal (p = 0.69) and lateral (p = 0.67) directions. Only vertical and longitudinal errors were reported as MV portal images were used for image verification. Chang et al observed improved setup accuracy with SGRT compared to tattoo/laser-based setup supported with a substantial reduction in the residual setup errors. The difference in the vector spatial deviation between both methods is statistically significant for SGRT setups (P < 0.05). A higher degree of variability is seen with tattoo-based setups.

      Efficiency

      Jimenez et al [
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ] reported a reduction in the average setup time from 8.0 ± 0.7 to 6.8 ± 0.8mins with SGRT for tattooless breast radiotherapy patient positioning. Similarly, in Ma et al [
      • Ma Z.
      • et al.
      Optical Surface Management System for Patient Positioning in Interfractional Breast Cancer Radiotherapy.
      ] the total time for setup, registration, and correction with CBCT (5.1 ± 0.2 min) is substantially reduced for SGRT alignment (1.1 ± 0.3 min). Batin et al [
      • Batin E.
      • et al.
      Can surface imaging improve the patient setup for proton postmastectomy chest wall irradiation?.
      ] achieved significant time savings in setup and treatment delivery (11 vs 6mins) for tattoo and SGRT setups, respectively. In contrast, Kost et al [
      • Kost S.
      • et al.
      Assessment of Setup Accuracy Using Anatomical Landmarks for Breast and Chest Wall Irradiation With Surface Guided Radiation Therapy.
      ] reported a slight increase in the average SGRT setup time from 5.4 to 6.3 min.

      Discussion

      This review aimed to determine the setup accuracy of SGRT compared to conventional tattoo/laser-based setup in breast radiotherapy and its correlation to IGRT (radiographic imaging). IGRT has evolved substantially over the past decade and encompasses various modalities that improve the quality of radiotherapy by minimising geometric uncertainties. SGRT is one such non-ionising surfacing imaging modality [
      • Sun B.
      • Chang J.
      • Rong Y.
      The more IGRT systems, the merrier?.
      ].
      The residual displacements post-IGRT were used for this systematic review. The hypothesis that SGRT is superior to conventional tattoo-based setups, is supported by the observed reductions in systematic, random errors, and the 3D vector displacements. SGRT improved the accuracy and reproducibility of patient setup compared to tattoo setups alone.

      Systematic and random errors

      The mean systematic errors in all translational directions were consistently reduced when compared to tattoo alignment. Systematic errors with tattoo setups ranged between 2.2 and 4.4 mm with average reductions of 0.4 to 2.5 mm, with SGRT. The mean systematic errors for SGRT were consistently below 2 mm, except for the longitudinal of 2.9 mm reported by Cravo et al [
      • Cravo Sá A.
      • et al.
      Radiotherapy setup displacements in breast cancer patients: 3D surface imaging experience.
      ]. Similarly, the mean random errors for SGRT were below 2 mm (except 2.7 mm for Cravo et al in the longitudinal direction). The mean longitudinal errors were larger for both tattoos and SGRT groups in the Cravo et al study [
      • Cravo Sá A.
      • et al.
      Radiotherapy setup displacements in breast cancer patients: 3D surface imaging experience.
      ] which could be predominantly attributed to chestwall motion as a result of breathing. Although the vertical and longitudinal mean systematic and random errors are reduced for the SGRT cohort, the impact of breathing motion is still present.
      The mean translational random errors indicate a greater degree of variability in the tattoo-only setup groups which ranged between 1.5 and 3.1 mm compared to 0.6–2.7 mm with SGRT. Significant reductions in systematic and random errors were reported by Batin et al [
      • Batin E.
      • et al.
      Can surface imaging improve the patient setup for proton postmastectomy chest wall irradiation?.
      ] in patients undergoing chestwall irradiation with proton therapy, with SGRT-aided initial alignment. The residual setup errors with SGRT were well within the 3 mm robustness for pencil beam scattering.
      The mean rotational errors were equivalent for tattoo and SGRT positioning. Traditionally the entire patient alignment is based on a few isolated tattoos/skin markings. However, SGRT provides a larger FOV, a ROI that encompasses the entirety of the treatment area and provides translational and rotational discrepancies. Rotational information can be used to physically manoeuvre and correct patient position prior to IGRT [
      • Padilla L.
      • et al.
      Assessment of interfractional variation of the breast surface following conventional patient positioning for whole-breast radiotherapy.
      ]. From the limited rotational error data available, both SGRT and tattoos provided good alignment to within 1degree. Larger reductions in the random rotational errors were seen in Batin et al [
      • Batin E.
      • et al.
      Can surface imaging improve the patient setup for proton postmastectomy chest wall irradiation?.
      ] and reductions of 0.1–0.4degrees across the other studies. Variations in the shape and size of the SGRT ROI between studies could potentially impact the surface registration results. Additionally, differences in setup procedure and IGRT registration methods were present between studies. The evaluated residual rotational errors from IGRT were based around the isocentre accounting for a limited subset of data and not the entire thorax and arm position.
      The systematic and random errors for SGRT in this review were maintained in the majority of cases sub 2 mm and 1degree in all translational and rotational directions. These improvements can be attributed to a larger FOV for postural correction and a ROI that incorporates patient topography and the target. During breast radiotherapy patients are typically immobilised on a breastboard or wingboard with one or both arms raised. Good postural alignment in breast radiotherapy is important as positional discrepancies of the arm can lead to larger interfraction setup deviations at the isocentre [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ,
      • Padilla L.
      • et al.
      Assessment of interfractional variation of the breast surface following conventional patient positioning for whole-breast radiotherapy.
      ,
      • Townend C.
      • et al.
      A review of permanent marking for radiotherapy in the UK.
      ,
      • Shah A.P.
      • et al.
      Clinical evaluation of interfractional variations for whole breast radiotherapy using 3-dimensional surface imaging.
      ]. Padilla et al [
      • Padilla L.
      • et al.
      Assessment of interfractional variation of the breast surface following conventional patient positioning for whole-breast radiotherapy.
      ] observed subsequent longitudinal shifts. Similarly, Kügele et al [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ] noted larger setup deviations in the tangential and locoregional patient groups resulting from incorrect placement of the patient arm. Visual feedback provided by SGRT, allows for postural correction (including the arm and chin) prior to IGRT and thereby minimises the residual errors. This cannot be achieved with tattoo/laser-based setups [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ]. Variability of tattoo positions due to skin elasticity and the need to manipulate the patient’s skin to match the lasers during the initial alignment, could be a contributing factor to the larger observed residual systematic setup errors. Tattoos are usually placed further away from the target with sole reliance on 2–6 points. The number of tattoos should not significantly influence the observed systematic errors.

      Mean residual setup errors

      This review found that the mean residual errors for SGRT (range: −2.1–2.3 mm) were consistently smaller than that of tattoo-only setups (-3.1 – 4.6 mm). There was a corresponding smaller variability in the distribution of the errors indicated by a smaller SD with SGRT. However, the residual errors in the vertical direction in 2 of the studies by Jimenez et al [
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ] and Kügele et al (Tangent group) [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ] appear larger for SGRT compared to tattoo/laser-based setups (2.3 vs −1.4 mm and 1.5 vs 0.6 mm respectively). Although a larger vertical error is observed in the SGRT group, Jimenez et al [
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ] suggest non-significant differences in the residual errors (5.9 vs 4.6 mm) between both setup methods based on surgical clip match on IGRT. The larger setup errors could be attributed to exacerbated chestwall motion with free-breathing techniques, compared to DIBH [
      • Cravo Sá A.
      • et al.
      Radiotherapy setup displacements in breast cancer patients: 3D surface imaging experience.
      ]. Additionally, the patient is potentially more relaxed at treatment compared to CT which is manifested in the vertical discrepancies and tattoos may be less sensitive to this occurrence [
      • Hattel S.H.
      • et al.
      Evaluation of setup and intrafraction motion for surface guided whole-breast cancer radiotherapy.
      ].
      Tattoo/laser-based setup as analysed on portal imaging is less sensitive to detecting soft tissue changes during the course of treatment, resulting from swelling or resolution of seromas, which could also explain this observation. Six studies reported that SGRT was sensitive in detecting errors in the vertical direction with smaller anterior-posterior discrepancies and may be beneficial for heart and lung dose sparing [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ,
      • Chang A.J.
      • et al.
      Video surface image guidance for external beam partial breast irradiation.
      ,
      • Liu M.
      • et al.
      Application of Optical Laser 3D Surface imaging system (Sentinel) in breast cancer radiotherapy.
      ,
      • Batin E.
      • et al.
      Can surface imaging improve the patient setup for proton postmastectomy chest wall irradiation?.
      ,
      • Cravo Sá A.
      • et al.
      Radiotherapy setup displacements in breast cancer patients: 3D surface imaging experience.
      ,
      • Shah A.P.
      • et al.
      Clinical evaluation of interfractional variations for whole breast radiotherapy using 3-dimensional surface imaging.
      ]. To account for breathing motion a new SGRT reference surface can be acquired post-IGRT and the user selects a surface that corresponds with the appropriate portion of the respiratory curve (e.g., mid-breathing). Alternatively, the SGRT system may average the patient breathing over a few seconds [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ,
      • Deantonio L.
      • et al.
      Detection of setup uncertainties with 3D surface registration system for conformal radiotherapy of breast cancer.
      ].

      3D vector displacements

      The results of this review indicate that the magnitude of the 3D vector shifts for SGRT was consistently reduced and in cases significantly less than the tattoo cohort, with reductions as large as 57% (14 vs 6 mm). 4 of the reviewed papers reported statistically significant reductions by a magnitude of 0.7 to 8 mm with SGRT compared to tattoos alone. Similarly, residual setup error reductions were reported in an additional 3 papers (range 0.3–1.3 mm) but were not statistically significant.
      Breast tissue is a non-rigid structure and challenging to position reproducibly; however, SGRT takes into consideration a larger area of patient surface topography and target, as opposed to just the tattoos, which explains the improved setup accuracy [
      • Hattel S.H.
      • et al.
      Evaluation of setup and intrafraction motion for surface guided whole-breast cancer radiotherapy.
      ,

      Shah C. et al., Evaluating radiotherapy options in breast cancer: does intraoperative radiotherapy represent the most cost-efficacious option? 2014;14:141-146.

      ]. Equally, variation in the arm position is detected and corrected thus minimising the influence of breast tissue deformation on isocentre accuracy [
      • Deantonio L.
      • et al.
      Detection of setup uncertainties with 3D surface registration system for conformal radiotherapy of breast cancer.
      ]. Similar improvements were seen with the use of SGRT across other clinical applications. Stanley et al [
      • Stanley D.N.
      • et al.
      Comparison of initial patient setup accuracy between surface imaging and three point localization: A retrospective analysis.
      ] observed statistically significant improvements in the 3D vector displacements with SGRT in a large sample of patients for various anatomical sites. Pelvis/lower extremities, abdomen, chest/upper extremities saw reductions in the magnitude of the 3D shift vector from 9 ± 4 mm to 6 ± 3 mm, 10 ± 5 mm to 5 ± 3 mm, and 9 ± 6 mm to 5 ± 3 mm, respectively.
      Skin markings are a useful surrogate for initial patient alignment however breast tissue is a superficial target that may not correspond directly to a bony match on the radiographs. MV portal images may still be utilised for verification in breast treatments, despite their limited image contrast, FOV, increased dose and challenge of interpreting the impact of rotations. The reliance on MV and KV imaging alone may result in underestimation of setup errors compared to CBCT and does not account for breast tissue deformation or reproducibility of the arm position [
      • Padilla L.
      • et al.
      Assessment of interfractional variation of the breast surface following conventional patient positioning for whole-breast radiotherapy.
      ,
      • Chang A.J.
      • et al.
      Video surface image guidance for external beam partial breast irradiation.
      ,
      • Laaksomaa M.
      • et al.
      AlignRT(®) and Catalyst™ in whole-breast radiotherapy with DIBH: Is IGRT still needed?.
      ,
      • Laaksomaa M.
      • et al.
      Estimation of optimal matching position for orthogonal kV setup images and minimal setup margins in radiotherapy of whole breast and lymph node areas.
      ]. SGRT is sensitive to these discrepancies and provides the user with additional postural and isocentric information [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ]. The use of a ROI that encompasses the entire breast and surrounding chestwall improves correlation to IGRT rather than bony anatomy match alone. This has been observed across a variety of treatment techniques [
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ,
      • Batin E.
      • et al.
      Can surface imaging improve the patient setup for proton postmastectomy chest wall irradiation?.
      ].

      Conformal radiotherapy techniques

      Jimenez et al [
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ], Chang et al [
      • Chang A.J.
      • et al.
      Video surface image guidance for external beam partial breast irradiation.
      ], and Hattel et al [
      • Hattel S.H.
      • et al.
      Evaluation of setup and intrafraction motion for surface guided whole-breast cancer radiotherapy.
      ] that included partial breast irradiation, or a subset of these patients, observed improvements in the setup accuracy and reproducibility with SGRT. Surgical clip matching on IGRT is often used as the ground truth for setup verification with these techniques. Whole breast radiotherapy may be more forgiving of small setup errors; however, partial breast irradiation includes the lumpectomy site and a surrounding margin, thus accurate target registration and setup reproducibility are crucial. Jimenez et al [
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ] reported a non-significant difference in the systematic and random errors between the tattoo and SGRT groups, suggesting that accurate patient setup can be achieved in APBI with SGRT and 2D surgical clip match, in the absence of any visible skin markings. This reinforces the non-reliance on tattoos in APBI under the above-mentioned conditions.
      Chang et al [
      • Chang A.J.
      • et al.
      Video surface image guidance for external beam partial breast irradiation.
      ] observed improved isocentric setup accuracy in partial breast irradiation with the use of SGRT, resulting in a significant reduction in the 3D vector shifts from 8.8 ± 4.2 mm to 4.0 ± 2.3 mm. An early study conducted by Bert et al [
      • Bert C.
      • et al.
      Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients.
      ] concluded a reduction in the mean 3D displacements and improved breast tissue congruence with SGRT compared to lasers and portal imaging alone in APBI. Similar results were seen by Gierga et al [
      • Gierga D.P.
      • et al.
      Comparison of Target Registration Errors for Multiple Image-Guided Techniques in Accelerated Partial Breast Irradiation.
      ] with good correspondence between target registration error for surgical clip match and SGRT to within 1 mm.
      The noted improvements in setup accuracy and reproducibility with SGRT provide greater confidence for tattooless breast radiotherapy, delivery of highly conformal treatments such as VMAT, partial breast irradiation, and more recently hypofractionated regimes [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ,
      • Leong B.
      • Padilla L.
      Impact of use of optical surface imaging on initial patient setup for stereotactic body radiotherapy treatments.
      ,
      • Chang A.J.
      • et al.
      Video surface image guidance for external beam partial breast irradiation.
      ,
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ,
      • Stanley D.N.
      • et al.
      Comparison of initial patient setup accuracy between surface imaging and three point localization: A retrospective analysis.
      ,
      • Batin E.
      • et al.
      Can surface imaging improve the patient setup for proton postmastectomy chest wall irradiation?.
      ,
      • Ma Z.
      • et al.
      Optical Surface Management System for Patient Positioning in Interfractional Breast Cancer Radiotherapy.
      ]. The latest publication by González-Sanchis et al [
      • González-Sanchis A.
      • et al.
      Surface-guided radiation therapy for breast cancer: more precise positioning.
      ] shows SGRT provides more accurate patient positioning and target localisation compared to skin marks, as evaluated on ExacTrac radiographic image registration of at least 3 surgical clips within the tumour bed.

      SGRT and IGRT correlation

      Depending on the anatomical site treated, SGRT could potentially reduce the frequency of routine IGRT if there is a good correlation between surface and internal anatomy [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ,
      • Freislederer P.
      • et al.
      Recent advanced in Surface Guided Radiation Therapy.
      ]. SGRT displacements are provided in relation to the treatment isocentre, thus calibration ensures coincidence between the optical and treatment isocentre to within 1 mm.
      The results of this review suggest a good correlation between SGRT and IGRT with residual errors of comparable magnitude for both methods or further reduced with SGRT. As seen in Ma et al [
      • Ma Z.
      • et al.
      Optical Surface Management System for Patient Positioning in Interfractional Breast Cancer Radiotherapy.
      ] the systematic errors were non-significant (p > 0.05) for SGRT and CBCT in vertical, longitudinal and lateral directions (0.62 ± 2.54, 0.18 ± 2.61, 0.49 ± 2.54 mm vs 0.65 ± 2.40, 0.40 ± 2.42, 0.41 ± 2.44 mm). Similarly, Deantonio et al [
      • Deantonio L.
      • et al.
      Detection of setup uncertainties with 3D surface registration system for conformal radiotherapy of breast cancer.
      ] reported errors only in vertical and longitudinal axes based on MV portal images and SGRT was non-inferior to tattoos (1.2 ± 2.6, 0.7 ± 1.7 mm vs 1.0 ± 0.8, 1 ± 1.3 mm). Significant improvements in the residual errors with SGRT compared to laser setup and KV orthogonal imaging (8.8 vs 4 mm, p = 0.02) were reported by Chang et al [
      • Chang A.J.
      • et al.
      Video surface image guidance for external beam partial breast irradiation.
      ]. SGRT has been noted to improve the accuracy and reproducibility of the initial patient positioning compared to tattoos/lasers alone due to the volume of surface data analysed [
      • Ma Z.
      • et al.
      Optical Surface Management System for Patient Positioning in Interfractional Breast Cancer Radiotherapy.
      ,
      • Shah A.P.
      • et al.
      Clinical evaluation of interfractional variations for whole breast radiotherapy using 3-dimensional surface imaging.
      ].
      Significant reductions in the setup deviations were reported by Kügele et al [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ] for SGRT based tangential and locoregional breast treatments. The agreement of SGRT based setups to IGRT within the 4 mm clinical tolerance was higher than that of laser-based setups in tangent treatments (95% vs 84%) and (70% vs 54%) for locoregional treatments. Similar improvements in setup reproducibility were observed by Stanley et al (p < 0.01) and Kost et al (p < 0.001) over tattoo/laser-based alignment [
      • Stanley D.N.
      • et al.
      Comparison of initial patient setup accuracy between surface imaging and three point localization: A retrospective analysis.
      ,
      • Kost S.
      • et al.
      Assessment of Setup Accuracy Using Anatomical Landmarks for Breast and Chest Wall Irradiation With Surface Guided Radiation Therapy.
      ]. Leong et al [
      • Leong B.
      • Padilla L.
      Impact of use of optical surface imaging on initial patient setup for stereotactic body radiotherapy treatments.
      ] indicated that KV imaging prior to CBCT in SABR treatments could be excluded in 80% of fractions with a maintained acceptance level of the CBCT registration. Similarly, Heinzerling et al [
      • Heinzerling J.H.
      • et al.
      Use of surface-guided radiation therapy in combination with IGRT for setup and intrafraction motion monitoring during stereotactic body radiation therapy treatments of the lung and abdomen.
      ] reported that SGRT replaced the pre-CBCT KV imaging and provided setup accuracy within 5 mm without the reliance on skin markings in thorax SABR treatments.
      The observed reductions in residual setup errors and incidence of rejected images with SGRT could allow for the minimisation of routine IGRT in breast radiotherapy. Batin et al [
      • Batin E.
      • et al.
      Can surface imaging improve the patient setup for proton postmastectomy chest wall irradiation?.
      ] highlighted residual errors with SGRT were well within the 3 mm and 2degree threshold and could replace orthogonal KV imaging with improved accuracy, efficiency, and reduction in imaging dose. An additional study by Batin et al [
      • Batin E.
      • et al.
      Reducing X-ray imaging for proton postmastectomy chest wall patients.
      ] suggested initial and weekly IGRT is sufficient for patient positioning and monitoring when SGRT is used daily.
      Consideration is given to the lessened impact from breathing motion with DIBH however Laaksomaa et al [
      • Laaksomaa M.
      • et al.
      AlignRT(®) and Catalyst™ in whole-breast radiotherapy with DIBH: Is IGRT still needed?.
      ] concluded IGRT is required on the 1st fraction and thereafter SGRT can be used for patient setup with a 3 mm isocentric accuracy, allowing a considerable reduction in IGRT. Weekly IGRT may still be required. Haraldsson et al [
      • Haraldsson A.
      • et al.
      Surface-guided tomotherapy improves positioning and reduces treatment time: A retrospective analysis of 16 835 treatment fractions.
      ] reported daily SGRT patient alignment with weekly MVCT in tomotherapy delivery in H&N, CNS, and thoracic cancers were feasible with the current target margins and a 4 min mean timesaving per fraction for thorax patients. Moser et al [
      • Moser T.
      • et al.
      Clinical evaluation of a laser surface scanning system in 120 patients for improving daily setup accuracy in fractionated radiation therapy.
      ] reported similar findings that SGRT improved the accuracy of initial patient setup and only intermittent imaging was required after the 1st fraction image verification.
      The reduction of residual setup errors with SGRT and its correspondence to IGRT makes it a suitable tool to reduce the frequency of imaging in breast radiotherapy and the corresponding patient imaging dose [
      • González-Sanchis A.
      • et al.
      Surface-guided radiation therapy for breast cancer: more precise positioning.
      ]. In departments where weekly IGRT verification is performed, SGRT facilitates accurate and reproducible daily alignment on non-imaging days [
      • Shah A.P.
      • et al.
      Clinical evaluation of interfractional variations for whole breast radiotherapy using 3-dimensional surface imaging.
      ].
      SGRT assumes a strong correlation between the surface and internal targets which is verified by IGRT [
      • Liu M.
      • et al.
      Application of Optical Laser 3D Surface imaging system (Sentinel) in breast cancer radiotherapy.
      ,
      • Batin E.
      • et al.
      Can surface imaging improve the patient setup for proton postmastectomy chest wall irradiation?.
      ]. Despite the improvement in the systematic and random translational errors, Cravo et al [
      • Cravo Sá A.
      • et al.
      Radiotherapy setup displacements in breast cancer patients: 3D surface imaging experience.
      ] suggest SGRT should be used with another imaging method due to its superficial characteristics. This sentiment is shared by Hattel et al [
      • Hattel S.H.
      • et al.
      Evaluation of setup and intrafraction motion for surface guided whole-breast cancer radiotherapy.
      ] where an improvement in the residual 3D vector displacements was observed for SGRT (5.4 to 4.2 mm), however, this was deemed not sufficient to replace the daily KV imaging.

      Efficiency

      SGRT improved setup accuracy as well as improved treatment efficiency, which could be attributed to redundancies in the tattoo/laser-based setups processes, reduced repeat imaging, or the replacement of daily or routine verification imaging [
      • Kügele M.
      • et al.
      Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
      ,
      • Jimenez R.B.
      • et al.
      Tattoo free setup for partial breast irradiation: A feasibility study.
      ,
      • Liu M.
      • et al.
      Application of Optical Laser 3D Surface imaging system (Sentinel) in breast cancer radiotherapy.
      ,
      • Batin E.
      • et al.
      Can surface imaging improve the patient setup for proton postmastectomy chest wall irradiation?.
      ,
      • Ma Z.
      • et al.
      Optical Surface Management System for Patient Positioning in Interfractional Breast Cancer Radiotherapy.
      ,
      • Deantonio L.
      • et al.
      Detection of setup uncertainties with 3D surface registration system for conformal radiotherapy of breast cancer.
      ,
      • González-Sanchis A.
      • et al.
      Surface-guided radiation therapy for breast cancer: more precise positioning.
      ,
      • Batin E.
      • et al.
      Reducing X-ray imaging for proton postmastectomy chest wall patients.
      ].

      Limitations of this review

      ROI delineation is important for SGRT surface registration and there were variations between the studies in this review. Chestwall patients may have a particularly uniform chest contour and the ROI should include suitable topography for accurate SGRT surface registration [
      • Ma Z.
      • et al.
      Optical Surface Management System for Patient Positioning in Interfractional Breast Cancer Radiotherapy.
      ]. This influence together with breathing motion may exacerbate the variation observed in the longitudinal direction [
      • Laaksomaa M.
      • et al.
      AlignRT(®) and Catalyst™ in whole-breast radiotherapy with DIBH: Is IGRT still needed?.
      ,
      • Batin E.
      • et al.
      Reducing X-ray imaging for proton postmastectomy chest wall patients.
      ]. Breath hold techniques could be beneficial in reducing interfractional and intrafractional variations [
      • Cravo Sá A.
      • et al.
      Radiotherapy setup displacements in breast cancer patients: 3D surface imaging experience.
      ]. Rigley et al [
      • Rigley J.
      • Robertson P.
      • Scattergood L.
      Radiotherapy without tattoos: Could this work?.
      ] showed comparable setup accuracy with right breast but statistically significant improvements for left breast cancers treated with DIBH and SGRT.
      The reference surfaces used varied between studies; however, Padilla et al [
      • Padilla L.
      • et al.
      Assessment of interfractional variation of the breast surface following conventional patient positioning for whole-breast radiotherapy.
      ] reported no significant differences between DICOM and SGRT reference surfaces. A survey conducted in the USA on the use of SGRT, revealed no distinct benefit of the SGRT over DICOM reference surfaces and vice versa [
      • Padilla L.
      • et al.
      A survey of surface imaging use in radiation oncology in the United States.
      ]. The DICOM reference is often preferred as it reproduces the initial planned position. In cases where the patient’s anatomy changes over a course of treatment, the SGRT reference surface may be useful.

      Conclusion

      For decades, skin marks and tattoos have been utilised in radiotherapy for initial patient alignment. The psychological implications of permanent skin markings are well documented. The evolution of SGRT with modern radiotherapy offers the ability to effect a change in clinical practice by negating the reliance on tattoos for patient setup in breast or chestwall treatments; whilst improving accuracy, patient experience, satisfaction, and quality of life. The results of this review suggest SGRT addresses the challenges associated with conventional tattoo-based alignment and augments the setup procedure providing a more accurate postural and isocentric setup. There appears to be a good correlation between SGRT and IGRT which could reduce the frequency of imaging and the associated patient dose.
      The aim of this study was not to compare the setup duration between both methods; however, there is evidence to suggest efficiency gains with SGRT which could positively impact the patient experience, radiotherapy staff, and the department alike. Continuous intrafraction motion management and automatic beam control with SGRT, allows for an additional level of safety with continually advancing radiotherapy delivery techniques.
      The trends observed in this review, further support the hypothesis that SGRT improves treatment setup accuracy without the permanent visual reminder of tattoos.

      Declaration of Competing Interest

      Wesley Naidoo is employed by VisionRT Ltd however this review was conducted independently with no influence or financial contribution from the employer.

      Appendix A. Supplementary data

      The following are the Supplementary data to this article:

      References

      1. Breast Cancer Now. Breast Cancer- The numbers. 2015. Available from: https://breastcancernow.org/sites/default/files/files/breast-cancer-stats-sheet-february-2015.pdf.

      2. National Institute for Health and Care Excellence (NICE). Early and locally advanced breast cancer: diagnosis and management- NICE guideline [NG101]. Available from: https://www.nice.org.uk/guidance/ng101/chapter/Recommendations#radiotherapy.

        • Koulis T.A.
        • Phan T.
        • Olivotto I.A.
        Hypofractionated whole breast radiotherapy: current perspectives.
        Breast cancer (Dove Medical Press). 2015; 7: 363-370
        • Murray Brunt A.
        • et al.
        Hypofractionated breast radiotherapy for 1 week versus 3 weeks (FAST-Forward): 5-year efficacy and late normal tissue effects results from a multicentre, non-inferiority, randomised, phase 3 trial.
        Lancet. 2020; 395: 1613-1626
        • De Felice F.
        • et al.
        Relation between Hypofractionated Radiotherapy, Toxicity and Outcome in Early Breast Cancer.
        Breast J. 2017; 23: 563-568
        • Coles C.E.
        • et al.
        International Guidelines on Radiation Therapy for Breast Cancer During the COVID-19 Pandemic.
        Clin Oncol. 2020; 32: 279-281
        • Laine A.M.
        • et al.
        The Role of Hypofractionated Radiation Therapy with Photons, Protons, and Heavy Ions for Treating Extracranial Lesions.
        Front Oncol. 2016; 5: 302
        • Michalski A.
        • et al.
        Inter- and intra-fraction motion during radiation therapy to the whole breast in the supine position: a systematic review.
        J Med Imaging Radiat Oncol. 2012; 56: 499-509
        • Townend C.
        • et al.
        A review of permanent marking for radiotherapy in the UK.
        Radiography. 2020; 26: 9-14
        • Pham K.
        • et al.
        Comparison of surface guided radiation therapy and tattooing techniques for breast radiotherapy.
        J Med Radiat Sci. 2019; 66: 147
        • Hoisak J.D.P.
        • Pawlicki T.
        The Role of Optical Surface Imaging Systems in Radiation Therapy.
        Semin Radiat Oncol. 2018; 28: 185-193
        • Kügele M.
        • et al.
        Surface guided radiotherapy (SGRT) improves breast cancer patient setup accuracy.
        J Appl Clin Med Phys. 2019; 20: 61-68
        • Nabavizadeh N.
        • et al.
        Image Guided Radiation Therapy (IGRT) Practice Patterns and IGRT's Impact on Workflow and Treatment Planning: Results From a National Survey of American Society for Radiation Oncology Members.
        Int J Radiat Oncol Biol Phys. 2016; 94: 850-857
        • Leong B.
        • Padilla L.
        Impact of use of optical surface imaging on initial patient setup for stereotactic body radiotherapy treatments.
        J Appl Clin Med Phys. 2019; 20: 149-158
        • Landeg S.J.
        • et al.
        A randomized control trial evaluating fluorescent ink versus dark ink tattoos for breast radiotherapy.
        Brit J Radiol. 2016; 89: 20160288
        • Padilla L.
        • et al.
        Assessment of interfractional variation of the breast surface following conventional patient positioning for whole-breast radiotherapy.
        J Appl Clin Med Phys. 2014; 15: 4921
        • Rigley J.
        • Robertson P.
        • Scattergood L.
        Radiotherapy without tattoos: Could this work?.
        Radiography. 2020; 26: 288-293
        • Moser T.
        • et al.
        Radiotherapy tattoos: Women's skin as a carrier of personal memory—What do we cause by tattooing our patients?.
        Breast J. 2020; 26: 316-318
        • Batista V.
        • et al.
        Clinical paradigms and challenges in surface guided radiation therapy: Where do we go from here?.
        Radiother Oncol. 2020;
        • Probst H.
        • et al.
        A systematic review of methods to immobilise breast tissue during adjuvant breast irradiation.
        Radiography. 2014; 20: 70-81
        • Hattel S.H.
        • et al.
        Evaluation of setup and intrafraction motion for surface guided whole-breast cancer radiotherapy.
        J Appl Clin Med Phys. 2019; 20: 39-44
        • Chang A.J.
        • et al.
        Video surface image guidance for external beam partial breast irradiation.
        Practical Radiation Oncol. 2012; 2: 97-105
        • Alderliesten T.
        • et al.
        Breast-shape changes during radiation therapy after breast-conserving surgery.
        Phys Imaging Radiation Oncol. 2018; 6: 71-76
        • Jimenez R.B.
        • et al.
        Tattoo free setup for partial breast irradiation: A feasibility study.
        J Appl Clin Med Phys. 2019; 20: 45-50
        • Stanley D.N.
        • et al.
        Comparison of initial patient setup accuracy between surface imaging and three point localization: A retrospective analysis.
        J Appl Clin Med Phys. 2017; 18: 58-61
        • Freislederer P.
        • et al.
        Recent advanced in Surface Guided Radiation Therapy.
        Radiation Oncol. 2020; 15
      3. Paxton JDPHAB, Waghorn B, Pawlicki T. Surface Guided Radiotherapy. 2020, Florida: CRC Press.

      4. Bekke SN. Evaluation of Optical Surface Scanning of Breast Cancer Patients for Improved Radiotherapy, in DTU Nutech. 2018. p. 175.

        • Liu M.
        • et al.
        Application of Optical Laser 3D Surface imaging system (Sentinel) in breast cancer radiotherapy.
        Sci Rep. 2020; 10: 7550
        • Wei X.
        • et al.
        Setup errors and effectiveness of Optical Laser 3D Surface imaging system (Sentinel) in postoperative radiotherapy of breast cancer.
        Sci Rep. 2018; 8: 7270
      5. PRISMA Transparent reporting of systematic reviews and meta-analyses. Welcome to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) website. 2021. Available from: http://www.prisma-statement.org/.

      6. McMaster University School of Nursing. EPHPP- Effective Public Health Practice Project. 2018. Available from: https://merst.ca/wp-content/uploads/2018/02/quality-assessment-tool_2010.pdf.

        • Batin E.
        • et al.
        Can surface imaging improve the patient setup for proton postmastectomy chest wall irradiation?.
        Pract Radiat Oncol. 2016; 6: e235-e241
        • Kost S.
        • et al.
        Assessment of Setup Accuracy Using Anatomical Landmarks for Breast and Chest Wall Irradiation With Surface Guided Radiation Therapy.
        Pract Radiat Oncol. 2019; 9: 239-247
        • Ma Z.
        • et al.
        Optical Surface Management System for Patient Positioning in Interfractional Breast Cancer Radiotherapy.
        Biomed Res Int. 2018; 2018: 6415497
        • Deantonio L.
        • et al.
        Detection of setup uncertainties with 3D surface registration system for conformal radiotherapy of breast cancer.
        Rep Pract Oncol Radiother. 2011; 16: 77-81
        • Sun B.
        • Chang J.
        • Rong Y.
        The more IGRT systems, the merrier?.
        J Appl Clin Medical Phys. 2017; 18: 7-11
        • Cravo Sá A.
        • et al.
        Radiotherapy setup displacements in breast cancer patients: 3D surface imaging experience.
        Rep Pract Oncol Radiother. 2018; 23: 61-67
        • Shah A.P.
        • et al.
        Clinical evaluation of interfractional variations for whole breast radiotherapy using 3-dimensional surface imaging.
        Practical Radiation Oncol. 2013; 3: 16-25
      7. Shah C. et al., Evaluating radiotherapy options in breast cancer: does intraoperative radiotherapy represent the most cost-efficacious option? 2014;14:141-146.

        • Laaksomaa M.
        • et al.
        AlignRT(®) and Catalyst™ in whole-breast radiotherapy with DIBH: Is IGRT still needed?.
        J Appl Clin Med Phys. 2019; 20: 97-104
        • Laaksomaa M.
        • et al.
        Estimation of optimal matching position for orthogonal kV setup images and minimal setup margins in radiotherapy of whole breast and lymph node areas.
        Rep. Practical Oncol Radiotherapy: J Greatpoland Cancer Center in Poznan and Polish Soc Radiation Oncol. 2014; 19: 369-375
        • Bert C.
        • et al.
        Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients.
        Int J Radiat Oncol Biol Phys. 2006; 64: 1265-1274
        • Gierga D.P.
        • et al.
        Comparison of Target Registration Errors for Multiple Image-Guided Techniques in Accelerated Partial Breast Irradiation.
        Int J Radiation Oncol*Biol*Phys. 2008; 70: 1239-1246
        • González-Sanchis A.
        • et al.
        Surface-guided radiation therapy for breast cancer: more precise positioning.
        Clin Transl Oncol. 2021;
        • Heinzerling J.H.
        • et al.
        Use of surface-guided radiation therapy in combination with IGRT for setup and intrafraction motion monitoring during stereotactic body radiation therapy treatments of the lung and abdomen.
        J Appl Clin Med Phys. 2020; 21: 48-55
        • Batin E.
        • et al.
        Reducing X-ray imaging for proton postmastectomy chest wall patients.
        Practical Radiation Oncol. 2018; 8: e266-e274
        • Haraldsson A.
        • et al.
        Surface-guided tomotherapy improves positioning and reduces treatment time: A retrospective analysis of 16 835 treatment fractions.
        J Appl Clin Med Phys. 2020; 21: 139-148
        • Moser T.
        • et al.
        Clinical evaluation of a laser surface scanning system in 120 patients for improving daily setup accuracy in fractionated radiation therapy.
        Int J Radiat Oncol Biol Phys. 2013; 85: 846-853
        • Padilla L.
        • et al.
        A survey of surface imaging use in radiation oncology in the United States.
        J Appl Clin Med Phys. 2019; 20: 70-77